CVE-2018-10634
Published: Aug 13, 2018
Modified: May 22, 2025
CVSS v3.1
4.8
Description
Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.
| Vendor | Product | Versions |
|---|---|---|
Medtronic | MMT- 508 - MiniMed pump | affected All versions |
Medtronic | MMT – 511 pump Paradigm | affected All versions |
Medtronic | MMT – 512 / MMT – 712 Paradigm x12 | affected All versions |
Medtronic | MMT – 515 / MMT – 715 Paradigm x15 | affected All versions |
Medtronic | MMT – 522 / MMT – 722 Paradigm REAL-TIME | affected All versions |
Medtronic | MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME | affected All versions |
Medtronic | MMT – 523 / MMT – 723 Paradigm Revel | affected All versions |
Medtronic | MMT – 523(K) / MMT – 723(K) Paradigm | affected All versions |
Medtronic | MMT – 554 / MMT – 754 MiniMed Veo | affected All versions |
Medtronic | MMT – 551 / MMT – 751 MiniMed 530G | affected All versions |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:N/A:N
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
Security Training
Train your team to recognize and prevent security threats with our comprehensive security awareness program.
Start TrainingVulnerability Scanning
Discover vulnerabilities in your applications and infrastructure before attackers do.
Scan Now