CVE-2018-14781
Published: Aug 13, 2018
Modified: May 22, 2025
CVSS v3.1
5.3
Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
| Vendor | Product | Versions |
|---|---|---|
Medtronic | MMT- 508 - MiniMed pump | affected All versions |
Medtronic | MMT – 511 pump Paradigm | affected All versions |
Medtronic | MMT – 512 / MMT – 712 Paradigm x12 | affected All versions |
Medtronic | MMT – 515 / MMT – 715 Paradigm x15 | affected All versions |
Medtronic | MMT – 522 / MMT – 722 Paradigm REAL-TIME | affected All versions |
Medtronic | MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME | affected All versions |
Medtronic | MMT – 523 / MMT – 723 Paradigm Revel | affected All versions |
Medtronic | MMT – 523(K) / MMT – 723(K) Paradigm | affected All versions |
Medtronic | MMT – 554 / MMT – 754 MiniMed Veo | affected All versions |
Medtronic | MMT – 551 / MMT – 751 MiniMed 530G | affected All versions |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N