Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

VendorProductVersions

Medtronic

24950 MyCareLink Monitor

affected
All versions

Medtronic

24952 MyCareLink Monitor

affected
All versions

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L

Attack Vector

Physical

Attack Complexity

High

Privileges Required

Low

User Interaction

None

Scope

Changed

Confidentiality

High

Integrity

Low

Availability

Low