CVE-2019-10964
Published: Jun 28, 2019
Modified: May 22, 2025
CVSS v3.1
7.1
Description
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
| Vendor | Product | Versions |
|---|---|---|
Medtronic | MiniMed 508 pump | affected All versions |
Medtronic | MiniMed Paradigm 511 pump | affected All versions |
Medtronic | MiniMed Paradigm 512/712 pumps | affected All versions |
Medtronic | MiniMed Paradigm 712E pump | affected All versions |
Medtronic | MiniMed Paradigm 515/715 pumps | affected All versions |
Medtronic | MiniMed Paradigm 522/722 pumps | affected All versions |
Medtronic | MiniMed Paradigm 522K/722K pumps | affected All versions |
Medtronic | MiniMed Paradigm 523/723 pumps | affected 0 - <= Software Versions 2.4A |
Medtronic | MiniMed Paradigm 523K/723K pumps | affected 0 - <= Software Versions 2.4A |
Medtronic | MiniMed Paradigm Veo 554/754 pumps | affected 0 - <= Software Versions 2.6A |
Medtronic | MiniMed Paradigm Veo 554CM/754CM pumps | affected 0 - <= Software Versions 2.7A |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability