Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

VendorProductVersions

Dräger

Perseus A500

affected
2.00
affected
2.01
affected
2.02
unaffected
2.03

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L

Attack Vector

Network

Attack Complexity

High

Privileges Required

None

User Interaction

None

Scope

Changed

Confidentiality

None

Integrity

None

Availability

Low