CVE Database
/

CVE-2019-25724

Back to search

CVE-2019-25724

Published: Jun 2, 2026

Modified: Jun 3, 2026

PUBLISHED

CVSS v3.1

6.5

MEDIUM

Description

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

VendorProductVersions

Dräger

Infinity M300

affected
VG2.x - <= VG2.3.1
unaffected
VG3.0

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Attack Vector

Adjacent

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Unchanged

Confidentiality

None

Integrity

None

Availability

High