CVE-2020-10627
Published: Dec 1, 2021
Modified: Aug 4, 2024
CVSS v3.1
7.3
Description
Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
| Vendor | Product | Versions |
|---|---|---|
Insulet | Omnipod Insulin Management System | affected 19191affected 40160affected ZXP425affected ZXR425 |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
References
Security Training
Train your team to recognize and prevent security threats with our comprehensive security awareness program.
Start TrainingVulnerability Scanning
Discover vulnerabilities in your applications and infrastructure before attackers do.
Scan Now