Back to search
CVE-2020-12045
Published: Jun 29, 2020
Modified: Aug 4, 2024
PUBLISHED
Description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
| Vendor | Product | Versions |
|---|---|---|
n/a | Baxter Sigma Spectrum Infusion Pumps | affected Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24 |
Weaknesses (CWE)
References
https://www.us-cert.gov/ics/advisories/icsma-20-170-04
x_refsource_MISC