Description

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

VendorProductVersions

Medtronic

Smart Model 25000 Patient Reader

affected
All versions

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H

Attack Vector

Adjacent

Attack Complexity

Low

Privileges Required

None

User Interaction

Required

Scope

Unchanged

Confidentiality

High

Integrity

High

Availability

High