CVE-2021-4479
Published: Jun 2, 2026
Modified: Jun 3, 2026
CVSS v3.1
4.0
Description
Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
| Vendor | Product | Versions |
|---|---|---|
Dräger | Atlan A350 | affected 0 - < 7.5.3affected 1.00 - < 1.01unaffected 2.0 |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability