Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

VendorProductVersions

Dräger

Atlan A350

affected
0 - < 7.5.3
affected
1.00 - < 1.01
unaffected
2.0

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L

Attack Vector

Network

Attack Complexity

High

Privileges Required

None

User Interaction

None

Scope

Changed

Confidentiality

None

Integrity

None

Availability

Low