CVE Database
/

CVE-2022-3027

Back to search

CVE-2022-3027

Published: Sep 13, 2022

Modified: Apr 16, 2025

PUBLISHED

CVSS v3.1

5.7

MEDIUM

Description

The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.

VendorProductVersions

Contec Health

CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor

affected
All

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:H

Attack Vector

Adjacent

Attack Complexity

Low

Privileges Required

None

User Interaction

Required

Scope

Unchanged

Confidentiality

None

Integrity

None

Availability

High