CVE Database
/

CVE-2024-48973

Back to search

CVE-2024-48973

Published: Nov 14, 2024

Modified: Nov 18, 2024

PUBLISHED

CVSS v3.1

9.3

CRITICAL

Description

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

VendorProductVersions

Baxter

Life2000 Ventilation System

affected
06.08.00.00 and prior

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Attack Vector

Local

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Changed

Confidentiality

High

Integrity

High

Availability

High

Security Training

Train your team to recognize and prevent security threats with our comprehensive security awareness program.

Start Training

Vulnerability Scanning

Discover vulnerabilities in your applications and infrastructure before attackers do.

Scan Now