CVE Database
/

CVE-2024-48974

Back to search

CVE-2024-48974

Published: Nov 14, 2024

Modified: Nov 18, 2024

PUBLISHED

CVSS v3.1

9.3

CRITICAL

Description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

VendorProductVersions

Baxter

Life2000 Ventilation System

affected
06.08.00.00 and prior

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Attack Vector

Local

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Changed

Confidentiality

High

Integrity

High

Availability

High