Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

VendorProductVersions

Baxter

Life2000 Ventilation System

affected
06.08.00.00 and prior

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Attack Vector

Local

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Changed

Confidentiality

High

Integrity

High

Availability

High