Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

VendorProductVersions

Contec Health

CMS8000 Patient Monitor

affected
All versions

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H

Attack Vector

Network

Attack Complexity

High

Privileges Required

None

User Interaction

Required

Scope

Unchanged

Confidentiality

High

Integrity

High

Availability

High