CVE-2025-0626
Published: Jan 30, 2025
Modified: Mar 1, 2025
CVSS v3.1
7.5
Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
| Vendor | Product | Versions |
|---|---|---|
Contec Health | CMS8000 Patient Monitor | affected All versions |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H