Back to search
CVE-2025-1204
Published: Feb 25, 2025
Modified: Feb 25, 2025
PUBLISHED
Description
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
| Vendor | Product | Versions |
|---|---|---|
Contec Health | CMS8000 Patient Monitor | affected 0 |