CVE Database
/

CVE-2025-14095

Back to search

CVE-2025-14095

Published: Dec 17, 2025

Modified: Dec 17, 2025

PUBLISHED

CVSS v3.1

6.8

MEDIUM

Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

VendorProductVersions

Radiometer Medical Aps

ABL90 FLEX and ABL90 FLEX PLUS Analyzers

affected
All application software versions with Windows 7, Windows XP as underlying OS
affected
Application software versions < 3.5MR11 with Windows 10 as underlying OS

Radiometer Medical Aps

AQT90 FLEX Analyzers

affected
All Application software versions <= 8.13 MR2

Radiometer Medical Aps

ABL800 BASIC and ABL800 FLEX Analyzers

affected
Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS
affected
Application software versions < 6.20 MR2 with Windows 10 as underlying OS

Radiometer Medical Aps

ABL9 Analyzers

affected
Application software versions < 1.5.0

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Attack Vector

Physical

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Unchanged

Confidentiality

High

Integrity

High

Availability

High