Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

VendorProductVersions

Radiometer Medical Aps

ABL90 FLEX and ABL90 FLEX PLUS Analyzers

affected
Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems
unaffected
Application software versions >= 3.5MR11 with Windows 10 operating system

Radiometer Medical Aps

ABL800 BASIC and ABL800 FLEX Analyzers

affected
Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems
unaffected
Application software versions >= 6.20MR2 with Windows 10 operating system

Radiometer Medical Aps

AQT90 FLEX Analyzers

affected
Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems
unaffected
Application software versions >= 8.13 MR2 with Windows 10 operating system

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

Attack Vector

Network

Attack Complexity

Low

Privileges Required

High

User Interaction

None

Scope

Unchanged

Confidentiality

High

Integrity

High

Availability

High