CVE-2025-15653
Published: Jun 2, 2026
Modified: Jun 3, 2026
CVSS v3.1
6.8
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
| Vendor | Product | Versions |
|---|---|---|
Dräger | Zeus IE | affected 0 - < 1.0.5 |
Dräger | Zeus RS C500 | affected 0 - < 1.0.9 |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability
Security Training
Train your team to recognize and prevent security threats with our comprehensive security awareness program.
Start TrainingVulnerability Scanning
Discover vulnerabilities in your applications and infrastructure before attackers do.
Scan Now