Published: Jun 2, 2026

Modified: Jun 3, 2026

PUBLISHED

CVSS v3.1

6.8

MEDIUM

Description

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

VendorProductVersions

Dräger

Zeus IE

affected
0 - < 1.0.5

Dräger

Zeus RS C500

affected
0 - < 1.0.9

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Attack Vector

Physical

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Unchanged

Confidentiality

High

Integrity

High

Availability

High