CVE-2025-24836
Published: Feb 13, 2025
Modified: Feb 14, 2025
CVSS v3.1
7.1
Description
With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.
| Vendor | Product | Versions |
|---|---|---|
Qardio | Heart Health IOS Mobile Application | affected 2.7.4 |
Qardio | Heart Health Android Mobile Application | affected 2.5.1 |
Qardio | QardioARM | affected All versions |
Weaknesses (CWE)
CVSS v3.1 Details
CVSS v3.1 Vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability