CVE Database
/

CVE-2022-43557

Back to search

CVE-2022-43557

Published: Dec 5, 2022

Modified: Apr 23, 2025

PUBLISHED

CVSS v3.1

5.3

MEDIUM

Description

The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.

VendorProductVersions

Becton, Dickinson and Company (BD)

BodyGuard™ Pump

affected
BD BodyGuard™
affected
CME BodyGuard™ 323 (2nd Edition)
affected
CME BodyGuard™ 323 Color Vision (2nd Edition)
affected
CME BodyGuard™ 323 Color Vision (3rd Edition)
affected
CME BodyGuard™ Twins (2nd Edition)

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H

Attack Vector

Physical

Attack Complexity

High

Privileges Required

None

User Interaction

None

Scope

Unchanged

Confidentiality

Low

Integrity

Low

Availability

High