CVE Database
/

CVE-2024-48967

Back to search

CVE-2024-48967

Published: Nov 14, 2024

Modified: Nov 15, 2024

PUBLISHED

CVSS v3.1

10.0

CRITICAL

Description

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

VendorProductVersions

Baxter

Life2000 Ventilation System

affected
06.08.00.00 and prior

Weaknesses (CWE)

CVSS v3.1 Details

CVSS v3.1 Vector

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Attack Vector

Network

Attack Complexity

Low

Privileges Required

None

User Interaction

None

Scope

Changed

Confidentiality

High

Integrity

High

Availability

High